Advantages of the industrial probe
- Measures relative humidity and temperature
- Wide temperature measurement range -100…200 °C
- Outstanding accuracy
- repeatability and long-term stability
- Advanced probe housing and construction
- Hot swappable, no calibration adjustment required due to the digital interface with Rotronic devices
Description
The Rotronic metal industrial probe is especially suitable for high temperatures up to 200 °C and demanding industrial environments and applications where hygiene plays an important role. It measures humidity and temperature and is compatible with the HF5, HF8 series measurement transmitters, HP32 and HP23 handheld devices and the HL-NT dataloggers.
Applications of the probe
- Food and pharmaceutical production
- Industrial Manufacturing
- Drying processes
Smart Electronics
-
- Based on the Rotronic’s AirChip3000
- Calculates the dew / frost point
- Alarm generation
- Saves adjustment data so that probes can be
interchanged without re-adjusting
- Can be cable mount up to 5 m
- Hot-swappable
Flexibility and Compatibility of the industrial probe
- User scalable analog ouput signals (2x 0…1V)
- Digital interface via UART
- Rapidly interfaced with HygroClip2 compatible devices from Rotronic or for OEM applications
Specification
| ACCURACY AT 23 ±5 °C | ±0.8 %rh / ±0.1 K |
|---|---|
| ACCURACY W. ADJUSTMENT PROFILE “STANDARD” | at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K |
| PROBE TYPE | Metal Industrial probe |
| OPERATING RANGE | -100…200 °C / 0…100 %rh |
| MEASURING RANGE | -100…200 °C (short-term peak load) / 0…100 %rh |
| POWER SUPPLY | 3.3…5 VDC |
| LONG-TERM STABILITY | <1 %rh / year |
| HUMIDITY SENSOR | Hygromer HT-1 |
| TEMPERATURE SENSOR | PT100 1/3 DIN Class B |
| PROBE PROTECTION | Filter carrier, brass nickel plated |
| FILTER CARTRIDGE | depending on the application, order separately |
| RESPONSE TIME T 63 | <15 sec. |
| ALLOWED WIND SPEED | 3.5 m/s, without filter |
| PSYCHROMETRIC CALCULATIONS | Dew or frost point |
| PROBE ADJUSTMENT BY SOFTWARE | 1 Point & multipoint %rh & °C |
| SOFTWARE UPDATE | With internet and HW4 software |
| USER INFORMATION | from Software HW4 (devicemanager) |
| ANALOGUE OUTPUTS SCALEABLE BY THE USER | Yes |
| ANALOG OUTPUT SIGNAL (STANDARD) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
| DIGITAL OUTPUT SIGNAL (STANDARD) | UART |
| TYPE OF OUTPUT SIGNALS | 2 x 0…1 VDC |
| MAX. CABLE LENGTH | Up to 5 m with passive cable. Up to 100 m with active cable. |
| PROBE MATERIAL | Stainless steel 1.4301 |
| REVERSE VOLTAGE PROTECTION | Mechanically in the connector |
| SERVICE INTERFACE | UART |
| FDA/GAMP COMPATIBILITY | FDA 21 CFR Part 11 and GAMP compatible |
| DIMENSIONS | HC2A-IM102-M: Ø15 x 86 mm x 2 m Cable; HC2A-IM302-M: Ø15 x 236 mm x 2 m Cable; HC2A-IM305-M: Ø15 x 236 mm x 5 m Cable |