Key features
- Pressure measurement from ±10 inH2O/2.5 kPa to 10000 psi/69 MPa
- Four absolute pressure measurement ranges: 15 PSIA (103 kPa), 30 PSIA (207 kPa), 100 PSIA (690 kPa), 300 PSIA (2 MPa)
- Pressure measurement accuracy 0.05% of full scale, 0.02% optional
- Current measurement range: 0-24mA
- Voltage measurement range: 0-30VDC
- 0.01% current and voltage accuracy
- Powerful HART communication, supports HART 7
- 24V loop power, no external power supply needed
- Lithium battery life: 25 hours continuously (without 24 V loop power), or 8 hours continuously (24 V loop power)
- Storage up to 128 sets of calibration data
- Ingress protection: IP54
- 1-meter drop test
- Three-year warranty
Description
Product overview: Fluke 730G Digital Pressure Calibrator
A compact, smart pressure solution for bench or field
The Fluke 730G Smart Digital Pressure Calibrator combines the performance of a precision test gauge with the power of a loop calibrator and also includes internal HART communication capabilities. When used with a pressure test pump the Fluke 730G becomes a high-efficiency, highly accurate calibrator that can test pressure gauges, pressure switches and pressure transmitters making it ideal for instrument shops, gauge labs and pressure laboratories.
The Fluke 730G Smart Digital Pressure Calibrator combines the performance of a precision test gauge with the power of a loop calibrator and also includes internal HART communication capabilities. When used with a pressure test pump the Fluke 730G becomes a high-efficiency, highly accurate calibrator that can test pressure gauges, pressure switches and pressure transmitters making it ideal for instrument shops, gauge labs and pressure laboratories.
- Built-in HART Communication: The integrated HART communication functionality is easy-to-use and allows you to adjust HART directly, without requiring additional tools for adjustment.
- Precision Pressure and Current Measurements: The 730G series provides high-precision 0.02% pressure measurement accuracy and 0.01% current measurement accuracy to meet the requirements for establishing a pressure laboratory and achieving accurate pressure calibration results.
- Intuitive operation and ease of use: A high-end LCD display and dual-jack design bring a more convenient and efficient user experience.
- Paperless testing and reporting: Optional 730G/PCSW management software enables paperless testing giving you the ability to automatically generate test reports and document results.
Fluke 730G/PCSW Software
- Paperless management of pressure calibrations: Record calibration data, automate calibration tests and generate calibration reports.
- Asset management: Manage tested equipment and pressure standards with ease.
- Supported pressure devices include: analog pressure gauges, digital pressure gauge, pressure transmitter, pressure sensors and pressure switches.
Increase your pressure testing capabilities using test pump kits
Fluke-700PTPK Pneumatic test pump kit:
- Combine with any Fluke-730G Series Gauge, 1000 psi (69 bar) or less
- Generate pressure up to 600 psi (40 bar) with the 700PTP-1 test pump
- Connect the 730G Series gauge directly to the included PTP-1 hand pump
- Includes hoses and adapters for connecting to the pressure device to be tested
- Hard case protects pump and gauge and allows gauge to remain connected to the pump
Fluke-700HTPK Hydraulic Test Pump Kits
- Combine with any Fluke-730G Series Gauge, 1000 psi (69 bar) or greater
- Generate pressure up to 10000 psi (690 bar) with the 700HTP-2 test pump
- Connect the 730G Series gauge directly to the HTP-2 hand pump
- Includes 700HTH-1 hoses and adapters for connecting to the pressure device to be tested.
- Hard case protects pump and gauge and allows gauge to remain connected to the pump
Specification
Specifications: Fluke 730G Digital Pressure Calibrator
| Technical specifications | ||||
| Pressure specifications: 0.02% of full scale (730RG), 0.05% of full scale (730G) | ||||
| Current specifications: 0.01% of reading | ||||
| Model | Range | Resolution | Accuracy1 | Burst pressure |
| 730G01 | -10 to +10 inH2O-20 to 20 mbar-2.49 to 2.49 kPa | 0.001 inH2O0.001 mbar0.0001 kPa | Positive pressure, ±0.2% of range2 Vacuum, ±0.2% of range2 |
3 psi,210 mbar20.6 kPa |
| 730G02 | -1 to +1 psi-70 to 70 mbar-6.89 to 6.89 kPa | 0.0001 psi0.001 mbar0.0001 kPa | 5 psi,350 mbar34.4 kPa | |
| 730G04 | -14 to 15 psi-0.97 to 1 bar-97 to 103 kPa | 0.001 psi0.0001 bar0.01 kPa | Positive pressure, ±0.05% of rangeVacuum, ±0.1% of range | 60 psi4 bar413.6 kPa |
| 730G05 | -14 to 30 psi-0.97 to 2 bar-97 to 206 kPa | 0.001 psi0.0001 bar0.01 kPa | 120 psi8 bar827.3 kPa | |
| 730G06 | -12 to 100 psi-0.83 to 6.9 bar-83 to 690 kPa | 0.01 psi, 0.0001 bar | 400 psi26 bar2.757 MPa | |
| 730G27 | -12 to 300 psi-0.83 to 20 bar-83 to 2 MPa | 0.01 psi0.001 bar0.1 kPa | 1200 psi80 bar8.273 MPa | |
| 730G07 | -12 to 500 psi-0.83 to 34 bar-83 to 3.4 MPa | 0.01 psi0.001 bar0.1 kPa | 2000 psi138 bar13.789 MPa | |
| 730G08 | -14 to 1000 psi,-0.97 to 69 bar-97 to 6900 kPa | 0.01 psi0.001 bar0.1 kPa | 4000 psi266 bar27.579 MPa | |
| 730G10 | -14 to 2000 psi,-0.97 to 140 bar-97 to 14 MPa | 0.1 psi0.01 bar1 kPa | 8000 psi550 bar55.158 MPa | |
| 730G29 | -14 to 3000 psi,-0.97 to 200 bar-97 to 20 MPa | 0.1 psi0.01 bar1 kPa | 10000 psi690 bar68.947 MPa | |
| 730G30 | -14 to 5000 psi,-0.97 to 340 bar-97 to 34 MPa | 0.1 psi0.01 bar1 kPa | 10000 psi690 bar68.947 MPa | |
| 730G31 | -14 to 10000 psi,-0.97 to 690 bar-97 to 69 MPa | 1 psi0.01 bar1 kPa | 15000 psi1035 bar103.421 MPa | |
| 730GA4 | 0 to 15 psia,0 to 1 bar0 to 100 kPa absolute pressure | 0.001 psi0.0001 bar0.01 kPa | ±0.05% of range4 | 60 psi4 bar413.6 kPa |
| 730GA5 | 0 to 30 psia0 to 2 bar0 to 200 kPa absolute pressure | 0.001 psi0.0001 bar0.01 kPa | 120 psi8 bar827.3 kPa | |
| 730GA6 | 0 to 100 psia,0 to 6.9 bar0 to 690 kPa absolute pressure | 0.01 psi0.0001 bar0.01 kPa | 400 psi27 bar2.757 MPa | |
| 730GA27 | 0 to 300 psia0 to 20 bar0 to 2 MPa absolute | 0.01 psi0.001 bar.01 kPa | 1200 psi80 bar8.273 MPa | |
| 730RG04 | -14 to 15 psi,-0.97 to 1 bar-97 to 100 kPa | 0.001 psi 0.0001 bar0.0001 kPa | Positive pressure, ±0.02 % of range3Vacuum, ±0.05 % of range3 | 45 psi3 bar310.2 kPa |
| 730RG05 | -14 to 30 psi,-0.97 to 2 bar-97 to 200 kPa | 0.001 psi 0.0001 bar0.01 kPa | 90 psi6 bar620.5 kPa | |
| 730RG06 | -12 to 100 psi,-0.83 to 6.9 bar-83 to 690 kPa | 0.01 psi 0.0001 bar0.01 kPa | 1000 psi69 bar6.894 MPa | |
| 730RG27 | –12 psi to 300 psi-0.83 to 20 bar–80 kPa to 2 MPa | 0.01 psi0.001 bar0.1 kPa | 2000 psi138 bar14 MPa | |
| 730RG07 | –12 psi to 500 psi-0.83 to 34 bar–80 kPa to 3.5 MPa | 0.01 psi0.001 bar0.1 kPa | 2000 psi138 bar13.789 MPa | |
| 730RG08 | -14 to 1000 psi,-0.97 to 69 bar-97 to 6900 kPa | 0.1 psi0.001 bar0.1 kPa | 10000 psi690 bar68.947 MPa | |
| 730RG29 | -14 to 3000 psi-0.97 to 200 bar-97 to 20 MPa | 0.1 psi0.01 bar1 kPa | 10000 psi690 bar68.947 MPa | |
| 730RG30 | -14 to 5000 psi,-0.97 to 340 bar-97 to 34 MPa | 0.1 psi0.01 bar1 kPa | 10000 psi690 bar68.947 MPa | |
| 730RG31 | -14 to 10000 psi,-0.97 to 690 bar-97 to 69 MPa | 0.1 psi0.01 bar1 kPa | 15000 psi1035 bar103.421 MPa | |
| 1 Accuracy specifications above have considered the effects of linearity, repeatability, resolution, reference accuracy, 1-year stability, and temperature effect.2 For 730G01/730G02, accuracy is ±0.3% of full scale @ 0 °C to 15 °C and 35 °C to 50 °C.3 For all 730RG model, add 0.003% FS/°C @ 0 °C to 18 °C and 28 °C to 50 °C。4 For 730GA4, add 0.006% FS/°C @ -10 °C to 15 °C and 35 °C to 55 °C; for all other 730GA model, add 0.003% FS/°C @ -10 °C to 15 °C and 35 °C to 55 °C. | ||||
| Media Compatibility730G01, 730G02, 730G04, 730G05, 730RG04, 730RG05: any clean dry non-corrosive gasAbove 100 psi: any liquids or gases compatible with 316 stainless steel | ||||
| Electrical specifications (All specifications are valid to 110% of range) | |
| Current | 0 mA to 24 mA |
| Voltage | 0 V DC to 30 V DC |
| Current range | 1 μA |
| Voltage range | 1 mV |
| Accuracy | 0.01% ± 2 LSD all ranges @ 23°C ±5 °C |
| Temperature compensation | From 20% FS/°C @ -10 °C to +18 °C and 28 °C to 50 °C |
| Loop Compliance Voltage | 24 V DC @ 20 mA |
| Temperature measurement accuracy inside the product | ±0.5 ℃ |
| Drive Capability | 1200 Ω (without HART resistor), 950 Ω (with internal HART resistor) |